SMG2000E Digital Display Clamp-on Phase Meter

SMG2000E Digital Display Clamp-on Phase Meter

Classification: Measurement Product/Secondary Circuit Test

Reference standard: JJG440-2008

Verification:  View the certificate

Instructions:  DOC Instructions

  PDF Instructions

Tel: 027-87207771 87207772 87207773 87188862

Verification certificate
Product introduction

SMG20000E Digital Display Clamp-on Phase Meter is special designed for voltage, current and phase tests, it is in high-precision, low price, portable hand-held, dual-channel input measuring. It's easy to measure the phase between U-U, I-I and U-I, identify the inductive circuit ,capacitive circuit and the phase sequence of three-phase voltage, detect the wiring group of transformer, test the secondary circuit and bus differential protection system, display the phase relationship among the CT groups of a differential protection system, and check if the wire of electric meter was correctly connected. When testing customers do not need cut off the tested current which input by clamp-current transformer. When measuring the phase between U1 and U2, the two input circuit loop are completely isolated, so the possible misconnection is absolutely avoided, which may result in short circuit and may even burn the meter. The monitor of the device is high-contrast LCD screen, and the character in big size, which helps to obtain the best visual effects.

Alternative name

Phase meters, double clamp phase meter, two-phase voltammetry table clamp, clamp phase meter, clamp phase voltammetry table, digital dual clamp phase meter, digital dual clamp phase meter, low voltammetry phase detection table, handheld digital dual phase voltammetry table clamp, double clamp digital phase meter, double clamp digital phase meter

The Characteristics of the products

1. Ingenious structure, easy to operate.
   a. hand held structure
   b. you should not disconnect the circuit or change the measurement range under 10mA-10A, 3V-500v.
   c. use high contrast LCD, character can be25mm, the screen angle can be 700, to obtain the best visual effects
   d. the function and layout of the switch are reasonable, rotating can read the measuring voltage ,current and phase.
2. High resolution
   adopt new type patent current clamp , current resolution up to 0.1mA; voltage resolution up to 0.1V.
3. Low power consumption
   this product with micro power consumption design and with the function of testing the voltage of battery.

The parameters of the products

Intrinsic Error
1. Reference Operating Conditions
  (1) Temperature:(23±5)℃
  (2) Moisture:(45~75)% RH
  (3) Wave forms of measured signal:sine wave、β=0.02
  (4) Frequency of measured signal:(50±0.2)Hz
  (5) Position of Measured current carrying conductor in the nipper jaw:Optional position
  (6) Amplitude range of measured signal when measuring phase:
      100~220V、0.5A~1.5A
  (7) External reference frequency electromagnetic interference: Should be avoided
2. Limits of Intrinsic Error
  (1) AC Voltage(see Table 1)


Table 1:Error of AC Voltage

Measure range

Resolution

Limits of intrinsic error

20V

0.01V

±(.3%RD+0.2% Range)

200V

0.1V

±(0.3%RD+0.2% Range)

500V

1V

±(0.3%RD+0.2% Range)

      Input Impedance:All the measure range is 2MΩ
      Voltage input impedance of phase testing: >500KΩ
  (2) Alternating current(see Table 2)


Table 2:Error of alternating current

Measure range

Resolution

Limits of intrinsic error

200mA

0.1mA

±(0.3%RD+0.2)

2A

1mA

10A

10mA

  (3) Phase
      U-U、U-I、I-I(see Table 3)


Table 3: Error of Power-frequency Phase

Range

Resolution

Limits of intrinsic error

0~360°

±2°

      Impedance of input voltage loop when measuring phase of U1-U2:40KΩ
Operating error
1. Rated Operating Conditions
  (1) Temperature:0℃~40℃
  (2) Moisture:≤90%v
  (3) Wave forms of measured signal:sine wave、β=0.05v
  (4) Frequency of measured signal:(50±0.5)Hz
  (5) Position of Measured current carrying conductor in the nipper jaw:Optional position.
  (6) Amplitude range of measured signal when measuring phase:v Phase U1-U2:30V~500V
      Phase I1-I2:10mA~10.00A
      Phase U1-I2 or I1-U2:10V~500V、10mA~10.00A
  (7) External reference frequency electromagnetic interference: should be avoidedv 2 . Limits of Rated Operating Error
    Under the conditions described in 1, the limits of the rated operating error will not exceed twice of the limits of intrinsic error.
Other technical features
1. Display:three and half
2. Sampling Rate:3 times per second
3. Power Supply:Single 9V laminated cell、current of power is less than 5mA
4. Dimension
   Meter shell: 186x86x33mm
   Nipper shell: 140mm×40mm×19mm
   Nipper jaw: Φ7mm×8mm
5. Weight
   Meter body: 280g
   Measuring pliers:2×200g
6. Storage Conditions
   Temperature:-10℃~50℃


Power supply: AC110V/115V/120V/127V/220V/230V/240V optional
Input frequency: 50Hz/60Hz optional
Note: The users can select the corresponding power which based on their local supply voltage and frequency, and we will confirm it before the delivery (or inform us that the goods will be used in which country advance)